Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

INFORMATION TO BE SUBMITTED UPON THE REGISTRATION OF DEVICES AND ECONOMIC OPERATORS IN ACCORDANCE WITH ARTICLES 26(3) AND 28, CORE DATA ELEMENTS TO BE PROVIDED TO THE UDI DATABASE TOGETHER WITH THE UDI-DI IN ACCORDANCE WITH ARTICLES 25 AND 26 AND THE UDI SYSTEM PART A INFORMATION TO BE SUBMITTED UPON THE REGISTRATION OF DEVICES AND ECONOMIC OPERATORS IN ACCORDANCE WITH ARTICLES 26(3) AND 28 Manufacturers or, when applicable, authorised representatives, and, when applicable, importers shall submit the information referred to in Section 1 and shall ensure that the information on their devices referred to in Section 2 is complete, correct and updated by the relevant party.
1. Information relating to the economic operator 1.1. type of economic operator (manufacturer, authorised representative, or importer), 1.2. name, address and contact details of the economic operator, 1.3. where submission of information is carried out by another person on behalf of any of the economic operators mentioned under Section 1.1, the name, address and contact details of that person,
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

1.4. name address and contact details of the person or persons responsible for regulatory compliance referred to in Article 15, 2. Information relating to the device 2.1. Basic UDI-DI, 2.2. type, number and expiry date of the certificate issued by the notified body and the name or identification number of that notified body and the link to the information that appears on the certificate and was entered by the notified body in the electronic system on notified bodies and certificates, 2.3. Member State in which the device shall or has been placed on the market in the Union, 2.4. in the case of class B, class C or class D devices: Member States where the device is or is to be made available, 2.5. presence of tissues, cells, or, their derivatives, of human origin (y/n), 2.6. presence of tissues, cells or their derivatives of animal origin as referred to in Regulation (EU) No 722/2012(y/n), 2.7. presence of cells or substances of microbial origin (y/n), 2.8. risk class of the device, 2.9. where applicable, the single identification number of the performance study,
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

2.10. in the case of devices designed and manufactured by another legal or natural person as referred in Article 10(14), the name, address and contact details of that legal or natural person, 2.11. in the case of class C or D devices, the summary of safety and performance, 2.12. status of the device (on the market, no longer placed on the market, recalled, field safety corrective Action initiated), 2.13. indication as to whether the device is a ‘new’ device. A device shall be considered to be ‘new’ if:
(a) there has been no such device continuously available on the Union market during the previous three years for the relevant analyte or other parameter;
(b) the procedure involves analytical technology not continuously used in connection with a given analyte or other parameter on the Union market during the previous three years.
2.14. indication as to whether the device is intended for self-testing or near-patient testing. PART B CORE DATA ELEMENTS TO BE PROVIDED TO THE UDI DATABASE TOGETHER WITH THE UDI-DI IN ACCORDANCE WITH ARTICLES 25 AND 26 The manufacturer shall provide to the UDI database the UDI-DI and the following information relating to the manufacturer and the device:
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

1. quantity per package configuration, 2. the Basic UDI-DI as referred to in Article 24(6) and any additional UDI-DIs, 3. the manner in which production of the device is controlled (expiry date or manufacturing date, lot number, serial number), 4. if applicable, the ‘unit of use’ UDI-DI (where a UDI is not labelled on the device at the level of its ‘unit of use’, a ‘unit of use’ UDI-DI shall be assigned so as to associate the use of a device with a patient), 5. name and address of the manufacturer, as indicated on the label, 6. the SRN issued in accordance with Article 28(2), 7. if applicable, name and address of the authorised representative (as indicated on the label), 8. the medical device nomenclature code as provided for in Article 23, 9. risk class of the device, 10. if applicable, name or trade name, 11. if applicable, device model, reference, or catalogue number, 12. additional product description (optional), 13. if applicable, storage and/or handling conditions (as indicated on the label or in the instructions for use), 14. if applicable, additional trade names of the device, 15. labelled as a single use device (y/n), 16. if applicable, the maximum number of reuses,
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

17. device labelled sterile (y/n), 18. need for sterilisation before use (y/n), 19. URL for additional information, such as electronic instructions for use (optional), 20. if applicable, critical warnings or contra-indications, 21. status of the device (on the market, no longer placed on the market, recalled, field safety action initiated). PART C THE UDI SYSTEM 1. Definitions Automatic identification and data capture (‘AIDC’) AIDC is a technology used to automatically capture data. AIDC technologies include bar codes, smart cards, biometrics and RFID. Basic UDI-DI The Basic UDI-DI is the primary identifier of a device model. It is the DI assigned at the level of the device unit of use. It is the main key for records in the UDI database and is referenced in relevant certificates and EU declarations of conformity. Unit of Use DI The Unit of Use DI serves to associate the use of a device with a patient in instances in which a UDI is not labelled on the individual device at the level of its unit of use, for example in the event of several units of the same device being packaged together. Configurable device
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

A configurable device is a device that consists of several components which can be assembled by the manufacturer in multiple configurations. Those individual components may be devices in themselves. Configuration Configuration is a combination of items of equipment, as specified by the manufacturer, that operate together as a device to achieve an intended purpose. The combination of items may be modified, adjusted or customised to meet specific needs. UDI-DI The UDI-DI is a unique numeric or alphanumeric code specific to a model of device and that is also used as the ‘access key’ to information stored in a UDI database. Human Readable Interpretation (HRI) HRI is a legible interpretation of the data characters encoded in the UDI carrier. Packaging levels Packaging levels means the various levels of device packaging that contain a fixed quantity of devices, such as a carton or case. Production Identifier (UDI-PI) The UDI-PI is a numeric or alphanumeric code that identifies the unit of device production. The different types of UDI-PI(s) include serial number, lot number, software identification and manufacturing or expiry date or both types of date.
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

Radio Frequency Identification (‘RFID’) RFID is a technology that uses communication through the use of radio waves to exchange data between a reader and an electronic tag attached to an object, for the purpose of identification. Shipping containers A shipping container is a container in relation to which traceability is controlled by a process specific to logistics systems. Unique Device Identifier (‘UDI’) The UDI is a series of numeric or alphanumeric characters that is created through a globally accepted device identification and coding standard. It allows the unambiguous identification of a specific device on the market. The UDI is comprised of the UDI-DI and the UDI-PI. The word ‘Unique’ does not imply serialisation of individual production units. UDI carrier The UDI carrier is the means of conveying the UDI by using AIDC and, if applicable, its HRI. UDI carriers include, inter alia, ID/linear bar code, 2D/Matrix bar code, RFID.
2. General requirements 2.1. The affixing of the UDI is an additional requirement — it does not replace any other marking or labelling requirements laid down in Annex I to this Regulation.
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

2.2. The manufacturer shall assign and maintain unique UDIs for its devices.
2.3. Only the manufacturer may place the UDI on the device or its packaging.
2.4. Only coding standards provided by issuing entities designated by the Commission pursuant to Article 24(2) may be used.
3. The UDI 3.1. A UDI shall be assigned to the device itself or its packaging. Higher levels of packaging shall have their own UDI.
3.2. Shipping containers shall be exempted from the requirement in Section 3.1. By way of example, a UDI shall not be required on a logistics unit; where a healthcare provider orders multiple devices using the UDI or model number of individual devices and the manufacturer places those devices in a container for shipping or to protect the individually packaged devices, the container (logistics unit) shall not be subject to UDI requirements.
3.3. The UDI shall contain two parts: a UDI-DI and a UDI-PI.
3.4. The UDI-DI shall be unique at each level of device packaging.
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

3.5. If a lot number, serial number, software identification or expiry date appears on the label, it shall be part of the UDI-PI. If there is also a manufacturing date on the label, it does not need to be included in the UDI-PI. If there is only a manufacturing date on the label, this shall be used as the UDI-PI.
3.6. Each component that is considered to be a device and is commercially available on its own shall be assigned a separate UDI unless the components are part of a configurable device that is marked with its own UDI.
3.7. Kits shall be assigned and bear their own UDI.
3.8. The manufacturer shall assign the UDI to a device following the relevant coding standard.
3.9. A new UDI-DI shall be required whenever there is a change that could lead to misidentification of the device and/or ambiguity in its traceability. In particular, any change of one of the following UDI database data elements shall require a new UDI-DI:
(a) Name or trade name, (b) device version or model, (c) labelled as single use, (d) packaged sterile, (e) need for sterilization before use, (f) quantity of devices provided in a package, (g) critical warnings or contra-indications.
Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance. )

article  annex_VI

CELEX:  02017R0746-20250110

3.10. Manufacturers that repackage or relabel devices with their own label shall retain a record of the original device manufacturer's UDI.
4. UDI carrier 4.1. The UDI carrier (AIDC and HRI representation of the UDI) shall be placed on the label and on all higher levels of device packaging. Higher levels do not include shipping containers.
4.2. In the event of there being significant space constraints on the unit of use packaging the UDI carrier may be placed on the next higher packaging level.
4.3. For single use class A and class B devices packaged and labelled individually, the UDI carrier shall not be required to appear on the packaging but it shall appear on a higher level of packaging e.g. a carton containing several packages. However, when the healthcare provider is not expected to have access, in cases such as in home healthcare settings, to the higher level of device packaging, the UDI shall be placed on the packaging.
4.4. For devices exclusively intended for retail point of sale, the UDI-PIs in AIDC shall not be required to appear on the point of sale packaging.