Directive 2006/7/EC of the European Parliament and of the Council of 15 February 2006 concerning the management of bathing water quality and repealing Directive 76/160/EEC article annex_V CELEX: 02006L0007-20140101 Rules on the handling of samples for microbiological analyses
1. Sampling point
Where possible, samples are to be taken 30 centimetres below the water's surface and in water that is at least one metre deep. 2. Sterilisation of sample bottles
Sample bottles are:
— to undergo sterilisation in an autoclave for at least 15 minutes at 121 oC, or
— to undergo dry sterilisation at between 160 oC and 170 oC for at least one hour, or
— to be irradiated sample containers obtained directly from manufacturer. 3. Sampling
The volume of the sampling bottle/container is to depend on the quantity of water needed for each parameter to be tested. The minimum content is generally 250 ml.
Sample containers are to be of transparent and non-coloured material (glass, polyethene or polypropylene).
In order to prevent accidental contamination of the sample, the sampler is to employ an aseptic technique to maintain the sterility of the sample bottles. There is no further need for sterile equipment (such as sterile surgical gloves or tongs or sample pole) if this is done properly.
The sample is to be clearly identified in indelible ink on the sample and on the sampling form. |
Directive 2006/7/EC of the European Parliament and of the Council of 15 February 2006 concerning the management of bathing water quality and repealing Directive 76/160/EEC article annex_V CELEX: 02006L0007-20140101 4. Storage and transport of samples before analysis
Water samples are to be protected at all stages of transport from exposure to light, in particular direct sunlight.
The sample is to be conserved at a temperature of around 4 oC, in a cool box or refrigerator (depending on climate) until arrival at the laboratory. If the transport to the laboratory is likely to take more than four hours, then transport in a refrigerator is required.
The time between sampling and analysis is to be kept as short as possible. It is recommended that samples be analysed on the same working day. If this is not possible for practical reasons, then the samples shall be processed within no more than 24 hours. In the meantime, they shall be stored in the dark and at a temperature of 4 oC ± 3 oC. |